6.9 Measuring
Waveform
6-85
6
How to Operate Sampling Scope
To set measurement items:
1. Click
Measure to open the Measure dialog box.
2. Click
Amplitude/Time tab.
3. Click on the right column of an item or items you want to add. The
column of the selected item turns into light blue.
4. Click
Add, and the light blue column is selected by
.
Click All Add Select Items, and the selected items are added to all
channels.
5. Click
the
Off button of Display to turn it On.
Make sure all the selected items are displayed in the Result window.
6. Click the Measure Setup button.
7. The item selected for the Item Selection list is added.
8. The measurement result is displayed in the button of the screen.
If the added item is the same as an item selected already for the same
channel, that item cannot be added using Item Selection.
To display measurement area
Set the range to measure One level and Zero level of the NRZ waveform
respectively.
1. Click
the
Setup (NRZ Amplitude Time) at Measure Setup. The Setup
(NRZ Amplitude Time) dialog box appears.
2. Click
Offset from Crossing and Width of EYE Boundary and set the
values respectively. Refer to “Figure 6.9.1.1-2 Setting Item of EYE
Boundary”.
Offset from Crossing
Width
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...