A.3 Sampling Oscilloscope
A-33
A
ppe
ndi
x A
A.3.12 26G/53G CRU (MP2110A-055)
Table A.3.12-1 Optical Data Input
Item
Specifications
Fiber
Singlemode fiber
Wavelength
1260 to 1620 nm
Data Format
NRZ, PAM4 available
Operation Baud Rate
51.0 to 58.0 Gbaud
25.5 to 28.9 Gbaud
Sensitivity (for 53G)
TypicalOuter OMA 100 µW*
1
Max. Outer OMA
265 µW*
2
Sensitivity (for 26G)
TypicalOuter OMA 100 µW*
3
Max. Outer OMA
265 µW*
4
Stress Signal
Sensitivity
Typical Outer OMA 630 µW*
5
Absolute Maximum
Power
+9.0 dBm (Average)
+12.0 dBm (Peak)
Optical Return Loss
Typical
–30 dB*
6
Tolerance against
consecutive 0s
500 bits or more at Zero Substitution Pattern PRBS 2
15
–1
Connector
FC/PC
*1: 53.125 Gbaud, PRBS13Q, PAM4, LBW = 4 MHz, TDECQ < 2.0 dB, at
25±5°C
*2: 51 Gaud, 53.125 Gbaud, 58 Gbaud, PRBS13Q, PAM4, TDECQ < 2.0
dB, LBW = 4MHz
*3: 26.5625 Gbaud, PRBS13Q, PAM4, LBW = 4 MHz, TDECQ < 2.0 dB,
at 25±5°C
*4: 25.5 Gbaud, 26.5625 Gbaud, 28.9 Gbaud, PRBS13Q, PAM4, TDECQ
< 2.0 dB, LBW = 4 MHz
*5: 53.125 Gbaud, SSPRQ, PAM4, LBW = 4 MHz, TDECQ
≈
3.4 dB, at
25±5°C
*6: At 1310 nm
Table A.3.12-2 Data Out SMF Optical
Item
Specifications
Insertion Loss
Typical
1.5 dB*
Max.
2.3 dB*
*: At 1310 nm
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...