5.2 PPG/ED
Panel
5-23
5
How to Op
erate BERT
Measurement method
The ED Result display can be refreshed either in real time (approx. 0.1 s
intervals) or when the progress bar reaches 100%. The display method is
set at Current of Gating.
On: Updates measurement results in real time.
Off: Updates measurement results when progress reaches 100% for either
Single or Repeat Gating Cycle.
Updates measurement results when measurement stops for Untimed
Gating Cycle.
The setting procedure of the measurement conditions is as follows:
1. Click the Gating Cycle button and select the measurement method as
follows:
Repeat
Single
Untimed
2. When Gating Cycle is Repeat or Single, click the text box under the
Gating Cycle and enter the numeric value.
The measurement cycle can be set from 1 second to 9 days 23 hours
59 minutes 59 seconds.
3. Click
Current to set the timing for the measurement result.
On:
Updates measurement results per 100 ms.
Off:
Displays measurement results per measurement results
or when measurement stops
4. Click
Start / Stop.
The right indicator of the button is changed to green.
Measure is displayed at the status display.
The measurement processing rate is displayed.
When pattern synchronization is obtained, the SYNC Loss display is
erased.
When Gating Cycle is set to Untimed, the measurement cycle progress is
displayed every 5 seconds.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...