6.9 Measuring
Waveform
6-107
6
How to Operate Sampling Scope
Jitter Analysis (Eye)
1. Click
Scope.
2. Click
Setup and set Sampling Mode to Eye. Set Signal Type to NRZ if
it is displayed.
3. Click
Time and set Data Clock Rate and Pattern Length.
4. Click
Sampling to display Sampling Run at the button display.
5. Click
Auto Scale.
6. To check NRZ? error, click Measure > Amplitude/Time tab.
7. Turn Display Result to On.
8. Check that the eye pattern is displayed at the center of screen and
NRZ? error is not displayed.
9. Set Item Selection to Item Type: Jitter.
10. To select an item or items, click on the right column.
11. Click Add. The measurement item is displayed on the Result window.
Figure 6.9.3-4 Display Example of Waveform
12. To change the measurement conditions, click Jitter of Measure
Setup.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...