Chapter 6 How to Operate Sampling Scope
6-56
Table 6.2.6-1 Items of Time Dialog Box (Cont’d)
Tab
Item
Description
CRU, CRU
(2
6G).
C
R
U
(53
G)
Clock Recovery
Sets the clock recovery unit when Operation Mode is Recovery.
Displayed only when Operation Mode is Recovery.
Lock Status
Whether or not the clock recovery unit is locked on the input signal
is indicated.
Operation
Rate
Selects the operating frequency of the clock recovery unit.
Specifies the symbol rate (kbaud) if the Operation Rate is Variable.
CRU Loop BW Selects the loop bandwidth of the clock recovery unit.
CRU Lock*
6
Allows the clock recovery unit to lock the recovered clock to the
input signal.
Auto Relock*
6
Off:
Does not perform CRU Lock automatically when a signal is
input to the clock recovery unit. To lock to the input signal,
click CRU Lock.
On:
Performs CRU Lock automatically when a signal is input to
the clock recovery unit.
Operation Clk
Rate
MP2110A-054:
When Operation Mode is Through, the frequency of the clock
input to the
CRU In
connector is displayed.
MP2110A-055:
When Operation Mode is Through, the frequency of the clock
input to the
Optical SMF Data In
connector is displayed.
*6: Displayed when MP2110A-055 is installed.
MP2110A-054 MP2110A-055
Green
The clock output from
CRU Out
is locked on the
input signal.
The clock output from
Recovered Clock Out
is
locked on the input signal.
Red
The clock output from
CRU Out
is NOT locked
on the input signal.
The clock output from
Recovered Clock Out
is
NOT locked on the input
signal.
No signal is input to
Optical SMF Data In
.
Orange
The clock output from
Recovered Clock Out
is
NOT locked on the input
signal.
A signal is input to
Optical SMF Data
In
but is
out of the symbol rate
range.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...