Chapter 6 How to Operate Sampling Scope
6-90
To set measurement method of rise/fall time
When measuring the rise time and fall time levels, select from amplitude
10/90% level or 20/80% level.
●
Click the Rise/Fall Time button to display 10/90% or 20/80%.
The bandwidth of the sampling oscilloscope is corrected with the
measurement value of the Rise/Fall time, and the corrected bandwidth is
displayed.
1. Click
NRZ at Measure Setup.
2. Click
Rise/Fall Time Correction to set the button display to On.
3. Click the Correction Factor text box.
4. Input the correction value in ps.
The calculation formula is as follows:
2
2
Tc
Tm
Td
Td: display value (ps), Tm: measurement value (ps),
Tc: correction value (ps)
If the measurement value is larger than the correction value, the result
displays N/A.
“*1” is displayed in red at the Rise/Fall time when Rise/Fall Time
Correction is set to On. When Rise/Fall Time Correction is On, “*1” is
displayed in red at Rise Time and Fall Time. “*1:Corrected” is displayed
in the measurement result display area and indicates that the values
with “*1” have been corrected. The number is subject to change from 1.
Figure 6.9.1.1-3 Display Example of Measurement Results
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...