6.9 Measuring
Waveform
6-101
6
How to Operate Sampling Scope
Adjusting mask position
To adjust mask position automatically
When Align Method is Zero/One/Crossing, click the Mask Alignment
Update.
When clicking the Mask Alignment Update calculate Zero/One/Crossing
of the currently described waveform and optimize the mask position
automatically.
The mask position is adjusted automatically if the Scale and Offset on the
screen are changed. However, the mask position cannot be changed.
To adjust mask position manually
When Align Method is User Defined, the mask position, width and
amplitude can be adjusted using the marker.
Figure 6.9.2-3 Example of Adjusting Mask Position Manually
1. Click the right side button of Alignment Marker and set Display On
to display the marker.
2. Click the X1 text box to set the mask position.
3. Click
the
X text box to set the mask width.
4. Click the Y1 text box to set the level 1.
5. Click
the
Y text box to set the level 0.
X1
X
Y1
Y
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...