5.2 PPG/ED
Panel
5-11
5
How to Op
erate BERT
PRBS
Pattern
Generation
Polynomial
Pattern
Length
Maximum
number of
contiguous 1s
Maximum
number of
contiguous 0
2^7–1 1+X
6
+X
7
127
7 6
2^9–1 1+X
5
+X
9
511
9 8
2^15–1 1+X
14
+X
15
32767
15 14
2^23–1 1+X
18
+X
23
8388607
23 22
2^31–1 1+X
28
+X
31
214748647
31 30
The ratio of 1s to the PRBS pattern length is 50%.
1/2 Clock Pattern
“1/2 Clock Pattern” means the repeating pattern of “1” and “0”.
The pattern of which the clock frequency is divided in half is output from
the
Data Out
connector and
Data
Out
connector.
Figure 5.2.3-2 Relationship between Data and Clock when 1/2 Clock
Pattern is set
1/16 Clock Pattern
“1/16 Clock Pattern” means the repeating pattern of “11111111” and
“00000000”.
The pattern of which the clock frequency is divided in one-sixteen is
output from the
Data Out
connector and
Data
Out
connector.
Figure 5.2.3-3 Relationship between Data and Clock when 1/16 Clock
Pattern is set
Clock
Data
1
1
0
1
0
1
0
1
0
0
Pattern
data
Time
Clock
Data
1 1 0 0 0 0 0 0 0 0 1 1 1 1 1 1 1 1 0 0 0
Pattern
data
Time
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...