Chapter 1 Outline
1-36
Linearity
It indicates the degree that the three Eye Amplitudes of the PAM4
waveform deviate from one third of the difference between Level 0 and
Level 3. If the three Eye Amplitudes are equal, the linearity is 1.
It is calculated with the smallest eye amplitude out of the three eye
amplitudes.
Example:
Eye Amplitude Upper: 32%
Eye Amplitude Middle: 36%
Eye Amplitude Lower: 31%
The Linearity in this case is
33.33
31
= 0.93
Mask Margin
This is the margin between the waveform and the eye mask in mask test
of eye pattern. The margin in the amplitude direction is the ratio to the
interval from the edge of the eye mask to the 1 Level or the 0 Level.
The margin in the time direction is the ratio to the time from the edge of
the eye mask to the cross point.
0%
50%
75%
Figure 1.5-20 Mask Margin
Optical Modulation Amplitude (OMA)
This indicates a difference at 1 level and 0 levels of the NRZ waveform.
It is the same as Eye Amplitude.
For PAM4 waveform, refer to “Outer OMA”.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...