Chapter 7 Performance Test
7-8
10. Click PPG/ED Ch1.
11. Set Amplitude to 0.3 Vp-p.
12. Click Scope. Configure the settings as shown below:
Dialog Box
Item
Setting Value
Measure -
Item
(A) Fall Time
Amplitude/Time
(A) Rise Time
(A)
Crossing
(A) Jitter RMS
13. Measure the amplitude, rise/fall time, and Data Crossing on the
sampling oscilloscope.
14. Connect the coaxial terminator to the
PPG1 Data Out
connector.
(Refer to Figure 7.1.3-4.)
15. Connect the
PPG1
Data
Out
connector to the
Ch A In
connector
of the sampling oscilloscope.
16. Repeat steps 4 to 13.
For PPG2 to PPG4, perform tests in the same way. For PPG3 and PPG4
test, change the Clock Output in step 4 to the following value.
Item
Setting Value
Clock Output
Ch3/4
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...