7.3 Performance Test for Sampling Oscilloscope
7-29
7
Performance T
est
With the MP2110A-021, test the amplitude accuracy for Ch B after
measuring the amplitude accuracy for Ch A.
1. Click
Scope. Configure the settings as shown below:
Dialog Box
Item
Setting Value
Amplitude
Ch B Scale
50 mV/Div.
Ch B Offset
50 mV
Ch B Attenuation
0 dB
Histogram Target
Channel
Channel
B
2. Set the DC voltage to 0 V.
3. Connect the DC Power Source to
Ch B In
connector.
Connect the center conductor of the coaxial cable to plus and the
shield to minus.
4. Click
Ch B to set to On.
5. Click
Sampling to set to Run.
6. Record a standard deviation (Std Dev) and an average value
(Mean) of the histogram measurement result.
7. Set the DC voltage to 200 mV.
8. Record the average value of the histogram measurement result.
9. Set the DC voltage to –200 mV.
10. Record the average voltage of the histogram measurement
result.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...