Chapter 6 How to Operate Sampling Scope
6-42
Table 6.2.4-7 Items of Jitter Measure Dialog Box - Advanced Tab
Item
Description
TJ Measurement
BER
Sets the bit error rate to measure the eye opening in the Bathtub graph.
Fixed RJ
Set Off to display the graph using RJ calculated from the measured
waveform.
Set On to display TJ graph by setting RJ an arbitrary value.
It is used to simulate the change of TJ by changing RJ value.
RJ Value
TJ is calculated with the use of the value entered here when Fixed RJ is
set to On.
Correction Factor
When setting the display to On, DJ (Scale), the correction coefficients of
RJ (Scale) and RJ (rms) can be entered. Corrected value(s) is(are)
indicated by “*1” in the measurement result display area.
DJ (Scale)
It is the correction coefficient of DJ.
The value calculated from the waveform multiplied by this number is
displayed in the measurement result.
Set to 1.00 for no correction.
RJ (Scale)
It is the correction coefficient of RJ.
The value calculated from the waveform multiplied by this number is
displayed in the measurement result.
Set to 1.00 for no correction.
RJ (rms)
RJ (d-d), RJ (rms) correction coefficient.
RJ is corrected with the following formula.
2
2
r
m
RJ
σ
m
:
Standard deviation of measured RJ
σ
r
: Correction Coefficient
RJ: RJ (d-d), RJ (rms) after correction.
When not correcting, set Correction Factor to Off.
Define Threshold
Sets the position detection method of the cross point relative to the
amplitude of the eye pattern.
Set it to Auto for automatic detection and Manual to specify the
position.
Manual Crossing
When Define Threshold is Manual, set the cross point position within
the range of 30 to 70% of the amplitude.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...