6.9 Measuring
Waveform
6-105
6
How to Operate Sampling Scope
Jitter analysis can measure different kinds of jitter at the same time.
Also, Eye Mask can be measured simultaneously.
Figure 6.9.3-2 Display Example of Jitter Analysis and Eye Mask Test
When Channel Math of Scope is Off, jitter can be analyzed without any
limit of pattern length.
When Channel Math of Scope is On, the jitter analysis can be performed
on the waveforms of pattern length up to 32768.
When Extinction Ratio Correction is set to On on the Amplitude, O/E
Dialog Box in Section 6.2.5, “*1” is displayed in red beside the corrected
measurement item.
When Fixed RJ is set to On on the Advanced tab of the Jitter Measure
dialog box (Figure 6.2.4-8), “*2” is displayed in red for RJ (d-d) and RJ
(rms).
When Sampling Mode is Advanced Jitter
Jitter analysis is executed for the waveform measured in Pulse mode of
Scope.
Jitter is measured for each point of bit rising and falling.
Therefore, TJ, Bathtub, RJ/PJ Histogram, DDJ Histogram, Composite
Histogram, Jitter Spectrum Display (PJ vs Frequency), and Jitter Display
for each bit (DDJ vs Bit or PDJ vs Bit) can be measured.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...