7.3 Performance Test for Sampling Oscilloscope
7-39
7
Performance T
est
Additive Jitter Test
1. Remove the 20 dB fixed attenuator from
PPG1 Data Out
connector. Refer to Figure 7.3.4-2 (a).
2. Connect a coaxial terminator to
PPG1 Data Out
connector.
3. Connect
the
Sync Out
connector and the
Trigger Clk In
connector using a coaxial cable.
4. Connect the Clk Out connector and the
Ch A In
connector o using
a coaxial cable.
5. Click
PPG/ED Ch1. Configure the settings as shown below:
Item
Setting Value
Reference Clock
Internal
Bit Rate
25781250 kbit/s
Sync Out
PPG1_1/8Clk
Clk Out
Ch1/2
Test Pattern (PPG)
PRBS2^31–1
PPG Data/XData
ON
6. Click
Scope. Configure the settings as shown below:
Dialog Box
Item
Setting Value
Setup Signal Type
NRZ
Sampling
Mode
Eye
Number
of
Samples
4050
Accumulation
Time
Persistency
Time
10.0
sec
Time - Rate
Tracking
Symbol Rate: PPG,
Divide Ratio:
UserDefined
Divide
Ratio
4
Time -
Unit
UI
Scale/Offset
UI on Screen
1 UI
Time - CRU
Operation Mode
Recovery
Operation
Rate
100GbE/4
(25.78125G)
CRU Loop BW
10 MHz
Amplitude Scale
10.0
mV/Div
Offset 0
mV
Histogram Histogram
On
Axis
Time
X1
0.00
UI
X2
1.00
UI
Y1
1
mV
Y2
–1
mV
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...