6.7 Collecting
Data
6-75
6
How to Operate Sampling Scope
In averaging, the mean value of the data count input at Averaging is
calculated and the results are displayed on the screen. However, when 1 is
input, no averaging is performed.
The number of averaging waveforms is displayed on the screen while
executing averaging.
The waveform is captured a hundred times when the following condition
is set, and a hundred averaging waveforms are displayed.
●
Accumulation Type
Average
●
Averaging 100
wfms
Figure 6.7-5 Example of Averaging Process
The averaging is calculated using the following formula.
Number of waveforms
≤
Averaging setting count:
Averaging setting count
≤
Number of waveforms:
Ave(n): Averaging value
S(n): Measurement value
M:
Averaging setting count
n: Number of waveforms
The size of the noise is inversely proportional to the square root of the
averaging setting count.
For example, if M is set to 100, the noise amplitude is compressed to
about 1/10 compared to when M is set to 1.
Ave 𝑛
𝑛
1
𝐴𝑣𝑒 𝑛
1
𝑆 𝑛
𝑛
Ave 𝑛
M
1
𝐴𝑣𝑒 𝑛
1
𝑆 𝑛
𝑀
Averaging 1
Averaging 100
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...