Chapter 6 How to Operate Sampling Scope
6-102
Clicking Center moves the marker without depending on the waveforms.
In this case, the setting values of X1,
Δ
X, Y1, and
Δ
Y are initialized.
Although the marker is not displayed at Display Off, the marker position
can be adjusted by changing the value in the text box.
To limit mask area
To examine in which part of the mask at the center the error occurs when
the error occurs in the mask test, the mask area can be limited.
Setting width and angle can limit the mask area.
If the mask area is limited, the error that occurs in an upper and lower
mask area is not measured.
Figure 6.9.2-4 Example of Limiting Mask Area
1. Click
Measure.
2. Click
Mask Test tab
3
Click the Target Channel button to set the channel.
4. Click the Mask Area Restriction button to set On.
5. Click
the
Angle text box to set the angle within the range of –90 to
90.
6. Click
the
Width text box to set the width within the range of 0.01 to
1.00.
Top
Bottom
Center
Angle 0
˚
Width 0.1 UI
Angle 45
˚
Width 0.1 UI
Angle 45˚
Width 0.3 UI
0.1 UI
0.3 UI
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...