Chapter 1 Outline
1-38
Outer OMA
It is the optical modulation amplitude of PAM 4 waveform, and it is the
difference between Level 3 and Level 0.
Overshoot
It is the ratio of the maximum power to Level 3 in the PAM4 waveform
and is calculated by the following formula:
Overshoot = Pmax –Level 3
Outer OMA
100
Pmax: Maximum power measured in the waveform
p-p (peak to peak)
Indicates difference between maximum and minimum values for signal
amplitude and data distribution width
For example, Vp-p means the difference between the maximum and
minimum values of the AC voltage.
Jitter p-p means the difference between the maximum and minimum
values of the jitter time distribution.
PRBS (Pseudo-Random Bit Sequence)
PRBS is the abbreviation for pseudorandom bit sequence. It is a bit string
approximating actual communications data with a random distribution of
“1” and “0”. In a bit length of 2
n
–1, n is called the number of the PRBS.
Using the BERTWave, n can be set to 7, 9, 15, 23, and 31.
Table 1.5-2 PRBS Bit Length
n
2
n
–1
7 127
9 511
15 32767
23 8388607
31 2147483647
The number of the PRBS used for measuring the Rx sensitivity is
determined by the communications standard.
RIN OMA
RIN OMA is a measurement of the ratio of the light source's noise power
to signal power (dB/Hz) based on OMA and is calculated by the following
formula.
RIN OMA = 10log
10
(RN
one
+RN
zero
)
2
OMA
2
×BW
(dB/Hz)
RNone:
Electrical random noise power measured at the top level
Use Level 1 for NRZ signals and Level 3 for PAM4 signals.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...