Chapter 6 How to Operate Sampling Scope
6-68
6.5.2
Setting Clock Rate and Divide Ratio
The symbol rate can be set by measuring the clock rate from the signal
input to the
Trigger Clk In
connector of the front panel.
1. Click
Time.
2. Click
the
Rate tab.
3. Click the Tracking button of the Data Clock Rate to set to Off.
4. Click the Recalculate Option button to select Symbol Rate.
5. Click
Acquire Clock Rate.
The frequency is displayed in the Clock Rate text box.
Confirm the waveform and signal level input to the
Trigger Clk In
connector when the frequency is not displayed.
6. When the clock frequency is not displayed at Clock Rate or the
displayed frequency is not correct, click the Clock Rate text box and
input the frequency.
7. Click the Divide Ratio text box and input the divide rate.
The symbol rate is calculated by the divide rate and clock rate.
When using the BERT Sync Out as an external clock, the divide ratio is
set as follows.
Sync Output settings
Pattern
Divide Ratio
1/8 Clock
–
8
1/16 Clock
–
16
1/40 Clock
–
40
Pattern Sync
PRBS 2^7–1
127
PRBS
2^9–1
511
PRBS
2^15–1
32767
PRBS
2^23–1
8388607
PRBS
2^31–1
214748647
When setting Sync Output to Pattern Sync, the data cannot be collected.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...