Chapter 5 How to Operate BERT
5-12
Pattern Logic
There are two types of logic; Positive Logic (POS) and Negative Logic
(NEG).
With positive logic, the voltage at the Data connector goes High when the
data is “1”.
With negative logic, the voltage at the Data connector goes Low when the
data is “1”.
Figure 5.2.3-4 Relationship between Logic Setting and Output Voltage
Waveform at Connector
Set the pattern using the following procedure.
1. Click the button for Test Pattern, and you will see the Test Pattern
dialog box.
2. Click the button for the pattern you want to set.
Data connector
voltage if POS
is set
Data connector
voltage if NEG
is set
Time
Time
0
0
Pattern data 0 1 0 0 1 1 0 1 0 1
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...