A.3 Sampling Oscilloscope
A-27
A
ppe
ndi
x A
Table A.3.10-2 Optical Channel Performance
Item
Specifications
Data Input
Optical Input Connector Ch A, Ch B, Ch C, Ch D
Connector
FC Connector (SMF Input, MMF Input)
Fiber
SMF Input:
62.5 µm GI Multimode fiber, accepts single mode fiber (When the installed
option is other than MP2110A-x30 or x40)
Singlemode fiber (When the installed option is MP2110A-x30 or x40)
MMF Input:
62.5 µm GI Multimode fiber, accepts single mode fiber
Wavelength
SMF Input:
860 to 1650 nm (When other than MP2110A-x30 and x40 is installed)
1260 to 1650 nm (When MP2110A-x30 or x40 is installed)
MMF Input:
800 to 860 nm
Bandwidth (No
Filter)
Typical 35 GHz (SMF Input)
Typical 25 GHz (MMF Input)
Filter Response
Supports the following specifications
<NRZ standards>
100GbE/4 (25.78125 Gbit/s)
100GbE/4 FEC (27.7393 Gbit/s)
OTU4 (27.952493 Gbit/s)
32GFC (28.05 Gbit/s)
<PAM4 standards>
400GbE/8 SMF (26.5625Gbaud 13.3GHz) [D]
400GbE/8 MMF (26.5625Gbaud 12.6GHz) [D]*
1
IEEE802.3cd Draft 2.0
400GbE/8 MMF (26.5625Gbaud 11.2GHz) [D]
400GbE/8 (26.5625Gbaud 19.34GHz) [D]
400GbE/4 SMF (53.125Gbaud 26.5625GHz)[D]
400GbE/4 SMF (53.125Gbaud 38.7GHz)[D]
64GFC SMF (28.9Gbaud 14.45GHz) [D]
64GFC MMF (28.9Gbaud 12.4GHz) [D]
*1: When MP2110A-095 is installed, Coherent Eye is set and Test
pattern is set other than Variable
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...