Chapter 5 How to Operate BERT
5-20
Threshold Level
Voltage level for evaluating “1” and “0”.
The input connectors are AC coupled, so set a voltage corresponding to the
signal waveform excluding the DC components.
The waveform excepting the waveform input to the connector and the DC
component when inputting the 1.2V LVCMOS signal to the
Data In
connector and
Data
In
connector is shown below.
Set the threshold level to the waveform excepting the DC component.
Figure 5.2.8-2 Waveform That Threshold Level Can be Set
Logic
Select the positive logic (POS) or negative logic (NEG).
Voltage (V)
Waveform without DC component
Time
Voltage (V)
Waveform input to
Data In
–
Data
തതതതതതത
In
connectors
Time
0
0
Threshold level
1
1
–
1
–
1
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...