Chapter 7 Performance Test
7-32
(3) Procedure
The SMF performance test for Ch B is explained below.
1. In case of Figure 7.3.3-1, connect the
Sync Out
connector and
the
Trigger Clk In
connector using a coaxial cable.
In case of Figure 7.3.3-2, connect the
Trigger Clk In
connector
and the signal generator
output
connector using a coaxial cable.
2. In case of Figure 7.3.3-1, click PPG/ED Ch1. Configure the
settings as shown below:
Item
Setting Value
Reference Clock
Internal
Bit Rate
Variable, 28 Gbit/s, 0 ppm
Sync Output
PPG_1/8 Clk
In case of Figure 7.3.3-2, set the signal generator as follows:
Frequency: 3.5
GHz
Amplitude: 0.5 Vp-p (at sine-wave –2.0 dBm)
3. Confirm that no light has been input to the optical connector
(SMF or MMF) for
Ch B In
.
4. Click
Scope. Configure the settings as shown below:
Dialog Box
Item
Setting Value
Setup Signal Type
NRZ
Sampling
Mode
Eye
Number
of
Samples
2048
Accumulation
Time Persistency
Limit Type
10.0 sec
Time Tracking Off
Divide
Ratio
8
Unit UI
UI On Screen
2
Software
Delay
0
Amplitude
Ch B Scale
20
W/Div.
Ch B Offset
60
W
Ch B Attenuation
0 dB
Input
Connector
(Wavelength)
SMF 1310 nm
Measure -
Display
On
Amplitude/Time Item
(Ch B) Average Power
(dBm)
5. Click
Calibrate Module of Amplitude. Click OK after completing
the calibration.
6. Prepare the variable optical attenuator for single-mode fiber.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...