Appendix D Performance Test Record Form
D-4
Table D.1-3 Pulse Pattern Generator Performance Test Record Form (With MP2110A-093)
Frequency Accuracy
Connector
Set Bit Rate
(kbit/s)
Minimum
Value (kHz)
Measured
Value (kHz)
Maximum
Value (kHz)
Pass/Fail
Sync Out
28200000
3524964.8
3525035.2 Pass/Fail
9500000
1187488.2
1187511.8 Pass/Fail
Measurement Channel PPG 1 2 3 4
Waveform (Amplitude)
Connector Set Amplitude
Minimum
Value
Measured
Value
Maximum
Value
Pass/Fail
Data Out
0.1 Vp-p
0.06 Vp-p
Vp-p
0.14 Vp-p Pass/Fail
0.8 Vp-p
0.62 Vp-p
Vp-p
0.98 Vp-p Pass/Fail
Out
0.1 Vp-p
0.06 Vp-p
Vp-p
0.14 Vp-p Pass/Fail
0.8 Vp-p
0.62 Vp-p
Vp-p
0.98 Vp-p Pass/Fail
Waveform
Connector
Item
Minimum
Value
Measured
Value
Maximum
Value
Pass/Fail
Data Out
Crossing
40%
%
60% Pass/Fail
Rise Time
ps
17 ps Pass/Fail
Fall Time
ps
17 ps Pass/Fail
Jitter (rms)
ps
0.9 ps Pass/Fail
Out
Crossing
40%
%
60% Pass/Fail
Rise Time
ps
17 ps Pass/Fail
Fall Time
ps
17 ps Pass/Fail
Jitter (rms)
ps
0.9 ps Pass/Fail
Skew
Connector Set Amplitude
Minimum
Value
Measured
Value
Maximum
Value
Pass/Fail
Data Out,
Out
0.3 Vp-p
ps
±8 ps Pass/Fail
Data
Data
Data
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...