Chapter 6 How to Operate Sampling Scope
6-30
Table 6.2.3-3 Items of Setup Dialog Box (Save tab)
Item
Description
EYE/Pulse Shot
Clicking Capture saves the screen file in the result
window.
Inverse Background
Color
Sets the color of the save screen at EYE/Pulse Shot.
Waveforms Only
On: Saves only the waveform displayed in the Result
window to a file. The jitter graph is not saved.
Off: Saves the whole Result window to a file. The jitter
graph is not saved.
Pattern Capture
Clicking Capture captures amplitude data of one cycle of
test pattern in the waveform being displayed and save it
in a test file. This function is available when Sampling
Mode is Coherent Eye and Test Pattern is other than
Variable.
The text file is saved to the following folder.
C:\Users\Public\Documents\Anritsu\MX210000A
\UserData\Result\TXT
Samples / UI
Sets the number of times that the amplitude is measured
per UI by Pattern Capture (number of samples).
Support Information
Clicking Save saves “BERTWaveSupportInformation.zip”
on the desk top. This file includes information that could
be useful for solving a problem. Attach the file when you
inquire us about the MP2110A behavior.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...