Chapter 6 How to Operate Sampling Scope
6-58
Calibration and Adjustment
6.3
6.3.1
Calibrating Level
The amplitude accuracy of the sampling oscilloscope is guaranteed after
performing calibration.
When the calibration is required, the error message “Calibration is
required” is displayed in the Setup dialog box in red.
Perform the calibration, when using the MP2110A for the first time or the
message is displayed.
The calibration procedures are as follows:
1. Check that no signals are input to the front-panel
Ch A In
,
Ch B In
, or
Trigger Clk In
connectors. If there are
Ch C
and
Ch D
, check that no
signals are input to these connectors.
2. Click
Setup to open the Setup dialog box.
3. Click
the
Utilities tab.
4. Click
Calibration, and then the message is displayed to confirm that
the signal is not input to the connector.
5. Click
OK.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...