Service Message
12-10
MS-860 Industrial Raster Reader User’s Manual
Service Message
Service Message Status
Service Message
Threshold
Resolution
Definition:
The service timer is reset on power-on. Thus the timer only records the
time that has elapsed since the last reset.
The message is sent every time the timer expires.
Definition:
When enabled, a message is sent whenever the system detects that a
user-set service time has expired.
Serial Cmd:
<
K409,status
,service message,threshold,resolution>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Serial Cmd:
<
K409,
status,
service message
,threshold,resolution>
Default:
SERVICE
Options:
Any 1 to 10 ASCII characters except
NULL < ,
or
>
.
Definition:
Sets the number of hours or minutes that will transpire before the Ser-
vice message is output.
Serial Cmd:
<
K409,
status,service message,
threshold
,resolution>
Default:
300
(seconds or minutes as set in
Resolution)
Options:
2 to 65,535
Definition:
Records time in seconds or minutes.
Serial Cmd:
<
K409,
status,service message,threshold,
resolution
>
Default:
Seconds
Options:
0 = Seconds
1 = Minutes