Calibration
MS-860 Industrial Raster Reader User’s Manual
4-7
Straight
Video Status
Scan Speed Status
Note: Do not calibrate motor speed if using the calibration database.
Definition:
The scan line is set as a centered straight line during calibration. This is
equivalent to Top Position = Bottom Position = 127 in the raster com-
mand
<K506,,127,127>
.
No permanent change is made to the raster novram configuration
parameters.
Definition:
Video status includes both
Gain and Tracking.
When checked,
Gain and Tracking settings
are fixed and is not calibrated.
When unchecked, they become part of the
calibration procedure.
Serial Cmd:
<
K521,
raster options,
video status
,scan speed status,laser power sta-
tus,laser framing status,symbology detect>
Default:
Enabled
Options:
0 = Disabled
1 = Enabled
Definition:
When checked,
Scan Speed is fixed and is
not calibrated.
When unchecked, it becomes part of the
calibration procedure.
Serial Cmd:
<
K521,
raster options,video status,
scan speed status
,laser power sta-
tus,laser framing status,symbology detect>
Default:
Enabled
Options:
0 = Disabled
1 = Enabled
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com