![Microscan MS-860 User Manual Download Page 64](http://html1.mh-extra.com/html/microscan/ms-860/ms-860_user-manual_1788323064.webp)
Communications
MS-860 Industrial Raster Reader User’s Manual
6-3
RS-232/422 Host Port
Includes host port connections and host protocols.
Host Port Connections
Baud Rate, Host Port
Parity, Host Port
Stop Bits, Host Port
Usage:
Can be used to transfer data faster or to match host port settings.
Definition:
The rate at which the reader and host transfer data back and forth.
Serial Cmd:
<
K100,baud rate
,parity,stop bits,data bits>
Default:
9600
Options:
0 = 600
1 = 1200
2 = 2400
3 = 4800
4 = 9600
5 = 19.2K
6 = 38.4K
7 = 57.6K
8 = 115.2K
Usage:
Only changed if necessary to match host setting.
Definition:
An error detection routine in which one data bit in each character is set to
1 or 0 so that the total number of 1 bits in the data field is even or odd.
Serial Cmd:
<
K100,
baud rate,
parity
,
stop bits,data bits>
Default:
Even
Options:
0 = None
1 = Even
2 = Odd
Usage:
Only changed if necessary to match host setting.
Definition:
One or two bits added to the end of each character to indicate the end of
the character.
Serial Cmd:
<
K100,
baud rate,parity,
stop bits
,
data bits>
Default:
One
Options:
0 = One
1 = Two
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com