Overview of Configuration Database
5-2
MS-860 Industrial Raster Reader User’s Manual
Configuration by ESP
Configuration by Serial Command
Overview of Configuration Database
Configuration Database Control
Database Setup
<
K253,
index,gain,AGC mode,tracking,raster top,raster bottom,raster
speed,laser framing status,laser on position,laser off position,laser
power,background color>
Database Operation
<
K252,
number of active indexes,
database cycles>
Database Mode
<
K254,
mode,count/time>
Save Current to Database
<
K253+,
database index>
Load Current from Database
<
K253-,
database index>
Definition:
Up to 10 configuration database indexes can be set up either manually
or derived in part from calibration results as explained in Chapter 4,
"Calibration"
.
When enabled, these settings will be applied sequentially
and repeatedly during the read cycle.
Usage:
Useful in applications in which a variety of symbol conditions require dif-
ferent settings that can be applied in real time. For example, a package
with a Code 39 and Code 128 symbol appearing consistently in different
locations as they move through the scan range during a read cycle can be
read by the reader if the proper settings are in the active configuration
database.