98
Version 2.1, Revision K
Daily Instrument Setup
6
Removing racks from the unload drawer
1. Press the button
at the unload drawer to open and remove the rack.
2. Press the button again after removal to close the unload drawer.
The status display for the unload drawer is updated after closing the drawer.
Note
•
After the oven step has ended, the mounting medium is not completely dry. Treat the slides
carefully during the removal from the rack to avoid a sliding of the coverglass.
•
Ensure that all racks are always removed from the unload drawer.
Warning
•
The unload drawer must be opened and all racks present must be removed – at the latest – when
the message is received indicating that the unload drawer is completely occupied
.
Failure to observe the message may lead to instrument faults and specimen impairment.
•
Finished coverslipped racks remain in the oven until all racks are removed from the unload drawer.
The longer dwell time in the oven can lead to specimen impairments.
•
If not all racks are removed from the unload drawer, this can lead to sample destruction.
•
To ensure a smooth operation of the instrument software, the instrument must be
restarted by the user at least every 3 days. This applies to both HistoCore SPECTRA ST and
HistoCore SPECTRA CV stand-alone units as well as in operation as a workstation. This also implies
the HistoCore SPECTRA CV must be restarted by the user at least every 3 days.
Fig. 76
•
The user receives an information message
if the unload drawer is open for longer than
60 seconds.