44
Version 2.1, Revision K
Operation
5
5.5 Displaying the drawers
The lower area of the main window
shows the status of the load drawer, the unload drawer
and the consumables (coverglass and mounting medium).
•
The stations depicted by a directional arrow in the instrument
symbolize the load
drawer.
•
The center area
displays the assigned and available positions in the unload drawer.
•
The right-hand area
shows the state of the filled consumables (coverglass and mounting
medium).
•
After closing the load drawers, the instrument automatically detects any inserted or removed racks.
•
The racks in both drawers are shown on the screen with the respective rack handle color
Available positions are displayed as empty.
1
2
3
Fig. 23
Note
The drawers can be opened when the drawer buttons
and
light up green.
A drawer button is illuminated in red and cannot be opened in the following cases:
•
If there is a rack transported by the instrument in the load or unload drawer,
•
If a rack is to be transferred from the HistoCore SPECTRA ST to the HistoCore SPECTRA CV,
•
If a rack is being turned in the rotator
•
If a rack is being inserted into the rack elevator or removed from the rack elevator.