71
HistoCore SPECTRA CV
Operation
5
1
2
3
4
5
Fig. 46
1
Coverglass size: 50 mm
2
Slides
3
Edge position of the coverglass on the slide
4
Position on the slide, coverglass size of 50 mm
5
Label field of the slide
5.9.5 Adjustment of the application volume
The HistoCore SPECTRA CV has a volume setting for the application volume of the mounting medium that
is set by default and validated by Leica. The default values are defined as
0
(=default) in the setting for
the parameter set
. These settings can be used immediately.
Note
Leica recommends checking the 0 value under the given conditions and requirements (e.g. specimen
size, specimen type and specimen thickness, temperature and humidity) in the lab and adjusting it
where appropriate.
1. To adjust the application volume, select the corresponding parameter set in the
Parameter sets
menu and press the
Edit
button
.
the application volume can be lowered or increased by
turning the roller
3. The menu also offers the option of changing an assigned color
or to assign a new
P. 67 – 5.9.2 Assigning a parameter set to a rack handle color)
4. Apply the settings you have entered by pressing the
Save
button
or press
Cancel
to return to the
Parameter sets
menu without saving.