Malfunctions and Troubleshooting
8
2
1
Fig. 112
Note
If the error cannot be remedied by the user using the specified procedure, notify a Leica service
technician and remove the coverslip line in question from service for the time being.
8.3.2 Removing a rack from the elevator of the coverslip line
Warning
•
If, during operation, a malfunction occurs in one of the two coverslip lines, the user will receive
information messages from the instrument, which will provide the user with instructions for
the potentially necessary removal of a rack from the elevator. Paying attention to the displayed
messages is mandatory.
• Always read the messages on the display first, then perform the task(s) described and then confirm
the messages (if applicable) to continue with the rescue process.
• Risk of injury! Always wear personal protective clothing (lab coat, cut-resistant gloves and safety
goggles)!
•
Specimens may become dried out during the interruption.
1.
Read and confirm the messages, then open the hood.
2.
Observe the following message. Before confirming, execute the tasks stated in the message.
Take the needle of the respective coverslip line from its holder laterally
and
insert it into the parking location
.