86
Version 2.1, Revision K
Daily Instrument Setup
6
6.3.4 Emptying the waste tray
Note
The HistoCore SPECTRA CV automatically detects defective/broken coverglass and places it in the
waste tray.
Warning
Wear cut-resistant safety gloves
P. 18 – 3.1 Standard delivery – packing list)
waste tray!
1. Open the hood.
2. Check the waste tray for broken coverglasses.
3. If broken coverglass exist, remove the waste tray
and clean it.
4. Then insert the waste tray back in the position provided for it
.
1
2
Fig. 65
5. Close the hood again.
6.3.5 Inspect Pick&Place module
1. Open the hood.
2. Check skids
, suction cups front and rear
and coverglass sensor
pin
for dirt and clean if necessary
P. 114 – 7.2.10 Cleaning the Pick&Place
.
3. Close the hood again.