114
Version 2.1, Revision K
Cleaning and Maintenance
7
7.2.10 Cleaning the Pick&Place module
•
For handling the Pick&Place module, see
P. 86 – 6.3.5 Inspect Pick&Place module)
.
Clean the following areas of the Pick&Place module with a lint-free cloth soaked with xylene:
1. Check the front and rear suction cups
at the Pick&Place module for dirt and
damages. Remove dirt with a lint-free cloth and a cloth soaked in xylene. Deformed or damaged
suction cups must be replaced
P. 114 – 7.2.11 Exchange suction cups)
2. Carefully clean the skids
on the left and right on the Pick&Place module from dirt.
3. Control the coverglass sensor pin
for glued-on materials and mobility.
L
Dispose of glass shards, splinters and dust in accordance with laboratory specifications.
Note
At the bottom of the Pick&Place module is the coverglass sensor pin between both suction cups
. The coverglass sensor pin can be moved freely if you can move it up and down by
touching it lightly with your fingertip.
1
2
3
2
Fig. 91
7.2.11 Exchange suction cups
1. Unplug the deformed and/or dirty suction cups from the Pick&Place module and dispose in
accordance with laboratory regulations.
2. Remove new suction cups from the package.
3. Attach the suction cups at the front
and rear
of the Pick&Place
module using a bent tweezers
.