Daily Instrument Setup
6
Note
If the oven step has been disabled or if the oven is completely shut down, the rack is transported – by
the gripper – directly from the elevator into the unload drawer. Proceed carefully while removing out of
the unload drawer, since the mounting medium is not yet completely dry and in the event of improper
removal, coverglass may shift out of place.
6.6 Starting the coverslipping operation
Note
Once initialization is complete and the preparations have been made
before starting the coverslipping operation)
the coverslipping operation can be started by inserting a
filled rack. It is important to make sure that the rack handle color matches the color of the parameter
sets that is suitable for starting and being carried out.
Warning
•
The rack can be inserted only using the load drawer. Accidentally inserting a rack into the unload
drawer can lead to a collision and thus to an instrument fault and a possible loss of specimens!
• Direct insertion in the instrument is not possible!
• Exercise caution when opening or closing the drawers! Crushing hazard! The drawers are motorized
and open or close automatically at the press of a button. Do not block the extension range of the
drawers.
1. Fold the handle of the rack into the upright position
2. If the load drawer
button is green, press it and open the drawer.
3. Insert the rack in an available position of the load drawer
.
4. Insert the rack so that both the
Leica
logo
on the front side of the rack and the
Front
on the top of the colored handle are pointing toward the user. The arrow
on top of the colored handle must point into the instrument.
1
2
3
4
Fig. 72