Version 2.1, Revision K
Daily Instrument Setup
6
1
2
Fig. 67
6.
If the filling level is below the minimum mark, the reagent vessel has to be refilled with the same
reagent
P. 32 – 4.6 Refilling consumables)
Note
• Always fill reagent vessels with xylene. Blending with other reagents shall be avoided.
• Cover the reagent vessels at the end of the work day to prevent evaporation of the filled reagent.
6.3.7 Unload drawer
1. Press the drawer button on the unload drawer
.
2. The unload drawer opens.
3. Remove any racks that may be present in the unload drawer.
4. Perform a visual inspection of the drawer inserts for dirt and clean them if necessary
P. 106 – 7.2.3 Load and unload drawers)
.
5. Then close the drawer by pressing the button.
Warning
Wear cut-resistant safety gloves
P. 18 – 3.1 Standard delivery – packing list)
unload drawer!