81
HistoCore SPECTRA CV
Daily Instrument Setup
6
9. Close the hood.
10. The mounting medium bottles that are reinserted are detected by the instrument software and
the
Module Status
Note
The CMS detects that a new mounting medium bottle has been inserted, and informs the user that
an Extended Prime is required for the coverslip line
L2
P. 122 – 7.3.2 Extended Prime)
means, air and any old mounting medium that is possibly still present is ousted from the hose system.
Only afterwards is the instrument ready to start.
6.3.2
Monitoring and refilling of the needle cleaning container
Note
The HistoCore SPECTRA CV has an automatic fill level scan mechanism in the needle cleaning
container. A warning message informs the user if there is not a sufficient quantity of xylene left in
the glass vessel. Despite the automated fill level scan, Leica recommends also carrying out a visual
inspection during daily startup to ensure that no refilling is necessary during daily routine work.
Warning
P. 32 – 4.6 Refilling consumables)
must be observed when handling
solvents like xylene!
1. Select the
Module Status
menu in the main menu and press
the
Prime/Clean
button in the desired coverslip line.
2. The needle moves automatically to the maintenance position.
3. Open the hood.
4. Take the needle out of the holder sideways and place it in the prime bottle
5. Turn the knurled screw
clockwise to lift the needle cleaning container.
6. The needle cleaning container
is lifted upwards and can be removed
7. Dispose remaining xylene in the needle cleaning container according to the laboratory
regulations.
8.
Outside of the instrument, use a Pasteur pipette to fill xylene into the glass cylinder
up to the edge of the plastic cap (approx. 10 ml).
9. Following this, put the needle cleaning container back into the correct position within the
instrument once again and push it completely in.
10. Take the needle out of the prime bottle and reinsert it into the holder.
11. In the Maintenance menu, press the
Close
button.
12. Close the hood.
13. The needle is automatically immersed in the needle cleaning container.