133
HistoCore SPECTRA CV
Malfunctions and Troubleshooting
8
Problem/malfunction
Possible cause
Correction
Insufficient mounting
medium is applied to the
slides
•
Quantity of mounting medium is
set too low.
•
Adjust the application quantity
of the mounting medium in the
parameter set
Volume calibration
menu.
Too much mounting
medium is dispensed
on the slides, or there is
excess mounting medium
on the coverglass
•
Quantity of mounting medium is
set too high.
•
Adjust the application quantity
of the mounting medium in the
parameter set
Volume calibration
menu.
•
Mounting medium residue on the
shifter or the shifter tongue (also
below the shifter tongue).
•
Check the shifter and shifter
tongue (also below the shifter
tongue) for contamination, clean
with a lint-free cloth moistened in
xylene if appropriate
During application of the
mounting medium, the
specimen on the slide was
damaged
•
Needle height not set correctly.
•
Do not begin any further
coverslipping operations in the
respective coverslip line and notify
Leica Service. The needle height
can be changed and calibrated
only by Leica Service.
The Pick&Place module
allows coverglass to fall
•
Dirty or deformed suction cups
•
Check the suction cups on
the Pick&Place module for
contamination and deformation
. Clean the
suction cups or replace them with
.
•
Additionally, observe the procedure
•
coverglass in the magazine have
been moistened and can no
longer be removed individually.
•
Remove the cover slip magazine
and replace it with a new one
.
Summary of Contents for 14051454200