7.2.15
Changing the active carbon filter
Note
The active carbon filter installed in the instrument helps reduce the amount of reagent vapors in the
exhaust air. The life cycle of the filter can fluctuate strongly depending on intensity of use and the
reagent configuration of the instrument. Therefore, replace the active carbon filter regularly, but at least
every three months, and dispose of it properly in accordance with the laboratory regulations applicable
in the country of use.
1.
The user can access the activated carbon filter
from the front of the instrument
by opening the upper cover
2.
The activated carbon filter can be accessed without any tools and can be removed by pulling on
the pull tab.
3. Write the insertion date on the white adhesive label on the front side of the activated carbon
filter
).
4.
Insert the new activated carbon filter in such a way that the pull tab is visible and reachable even
after it is completely pushed in.
5.
The active carbon filter must be pushed in until you feel it make contact with the rear panel of
the instrument.
6. Close the cover