38
Version 2.1, Revision K
Installation and Instrument Setup
4
Correct fill level of the reagent vessels
Note
• When filling the reagent vessels, observe the fill level marks inside the reagent vessels.
•
Use the label covers contained in the
P. 18 – 3.1 Standard delivery – packing list)
to label the
reagent vessels in the load drawers.
•
The label covers with the letter
S
indicate to the user that the
reagent vessel has been filled with a solvent (xylene).
•
Use the non-labeled covers only in cases where the reagent vessels are empty (e.g. after shutting
down the instrument).
A sufficient fill level is ensured if the level of the reagents is between the maximum
and the
minimum
fill level mark.
Inserting the reagent vessels into the load drawer:
1. To insert the reagent vessels, press the drawer button for the load drawer
2. The load drawer opens.
3.
Remove the reagent vessel and fill it with a HistoCore SPECTRA ST xylene outside the instrument
in accordance with the safety instructions.
4.
Then insert the filled vessel back into the load drawer
5. Close the load drawer by pressing the drawer button again.
Fig. 18
1
2