36
Version 2.1, Revision K
Installation and Instrument Setup
4
1
2
4
4
3
3
3
6
5
Fig. 16
3. Carefully insert the cannula
with the label 1 into the opening of the mounting
medium bottle until it clicks into place. This process involves puncturing the white protective
membrane.
Warning
Puncturing the white protective membrane briefly requires an increase in force. Therefore, proceed
carefully to avoid slipping from your hand.
4. Repeat the process for bottle
L2
.
back until you feel it click into place.
6. Close the hood.
7. The mounting medium bottles that are inserted are detected by the instrument software and the
Module Status
is updated.
Note
The software identifies the insertion of the new bottles and informs the user that an
Extended Prime
is necessary for both coverslip lines
P. 122 – 7.3.2 Extended Prime)
. Only afterwards is the
instrument ready to start.