Cleaning and Maintenance
7
Note
The individual rinsing programs are clarified in the following, using the coverslip line
L1
The same procedures also apply for the coverslip line
L2
.
2
3
4
1
Fig. 98
7.3.1 Quick Prime
Note
At the start of operation, the user is prompted after starting the instrument to carry out the
Quick
Prime
rinsing program. This step is necessary to ensure the flow rate of the mounting
medium through the needle system. The respective coverslip line is ready to operate only after Quick
Prime.
1. After the steps under
have been completed, press the
Quick
Prime
button.
2. Priming lasts approx. 35 seconds and uses approx. 2 ml of mounting medium. This quantity is
taken into account by the CMS.
3. After the rinsing process has been completed, reinsert the needle unit into the holder
and confirm the corresponding information message with
OK
.