20
Version 2.1, Revision K
Instrument Components and Specifications
3
HistoCore SPECTRA CV
Dimensions and weights:
Dimensions
(length x depth x height):
Hood closed: 690 x 785 x 615 mm
Hood opened: 690 x 785 x 1060 mm
Weight, empty (without
reagents and accessories):
115 kg
Weight, filled (with reagents
and accessories):
120 kg
Workstation dimensions
and weights
(HistoCore SPECTRA CV and
HistoCore SPECTRA ST):
Dimensions
(length x depth x height):
Hood closed: 2044 x 785 x 615 mm
Hood opened: 2044 x 785 x 1060 mm
Weight, empty (without
reagents and accessories):
280 kg
Weight, filled (with reagents
and accessories):
335 kg
Performance:
Usable slides:
In accordance with DIN ISO 8037-1
(76 mm x 26 mm)
Coverglass cartridge capacity: Exclusively Leica consumables with
300 coverglasses per magazine
Coverglasses:
Exclusively Leica consumables.
Available size:
50 mm x 24 mm, thickness: No. 1
In accordance with ISO 8255-1
Mounting medium application
quantity:
Preset value corresponding to
coverglass size. Fine adjustment by
users possible.
Types of mounting media:
Exclusively Leica consumables:
X1 mounting medium
Capacity of mounting medium
bottle:
At least 1600 slides
Racks:
Leica rack for 30 slides
Factory settings:
Mounting medium application
volume:
Oven temperature:
40 °C (not changeable)
Oven step:
Enabled
Date format:
International (DD.MM.YYYY)
Time format:
24 h
Language:
English
Note
When using an external uninterruptible power supply (UPS), it should be designed for an output of at
least 1100 VA and ensure operation over a time frame of at least 10 minutes.