27
HistoCore SPECTRA CV
Installation and Instrument Setup
4
4.3 Exhaust air connection
» Connect one end of the exhaust air hose
on the top side of the instrument using the hose clamp
P. 18 – 3.1 Standard delivery –
included in the standard scope of delivery
Connect the other end to an
exhaust air device installed at the lab.
1
2
Fig. 7
Warning
•
A connection to an external exhaust air extraction system is highly recommended. The
active carbon filter must be used to enable the integrated air extraction system to reduce the
concentration of solvents in the air.
• The active carbon filter must be used for connecting to an external exhaust system as well
P. 117 – 7.2.15 Changing the active carbon filter)
. Compliance with this is mandatory.
•
The owner/operator must verify compliance with the workplace limit values when work is done with
hazardous materials.
4.4 Installing the accessories
4.4.1 Fit the drawer inserts into the unload drawer
Note
The load and unload drawers can be manually opened while the system's power is turned off.
1. Manually pull the unload drawer until it stops.
2.
For a simplified fitting of the inserts, the unload drawer can be further opened. For this purpose,
raise the red retaining lever
that moves the unload drawer completely out of the
instrument and carefully fold it downwards.
3. Insert the three inserts
into the compartments
in the unload drawer.