122
Version 2.1, Revision K
Cleaning and Maintenance
7
Note
The needle has a notch
that fits exactly in the holder. The Attention symbol
indicates to the user that utmost care is required when
inserting the needle into the holder. The needle must be inserted straight and all the way in order to
make sure that no negative impact on the samples arises during processing.
4. Repeat steps 1-3 for the second coverslip line (
L2
).
5. Finally, close the hood.
After checking the fill level, the instrument is ready to operate and can begin processing.
7.3.2 Extended Prime
Note
•
The
Extended Prime
rinsing program must be carried out after each time a mounting
medium bottle is replaced
P. 78 – 6.3.1 Changing the mounting medium bottle)
for the
respective coverslip line
L1
or
L2
.
•
During an Extended Prime process, mounting medium is pumped through the entire hose system to
ensure that no air remains in the hose system as a result of the replacement.
•
After the replacement of the mounting medium bottle, the
Quick Prime
and
Clean
buttons are inactive (grayed out).
1. To change the mounting medium bottle, proceed the same way as in
Changing the mounting medium bottle)
2. Take note of the information message that follows
Ok
button.
Fig. 99
3. After the steps under
have been completed, press the
Extended Prime
button.