146
Version 2.1, Revision K
Malfunctions and Troubleshooting
8
16. Follow the instructions on the display and close the locking mechanism of the elevator again
and push the elevator a few centimeters downward
. Confirm the
message after completing the task.
1
2
Fig. 119
17. The next message asks the user to remove the needle from the parking location
and insert it into the holder
of the coverslip line. Make sure that the needle is
positioned correctly and confirm the message.
2
1
Fig. 120
18. Close the hood.
19. After the hood has been closed, a new message appears informing the user to check for
uncoverslipped slides
in the removed rack, considering the coverslipping
. Uncoverslipped slides have to be coverslipped immediately to avoid
drying out of the specimen. Confirm the message.