Malfunctions and Troubleshooting
8
8.3.1 Malfunction at the coverglass receptacle
If the
Pick&Place module
loses a coverglass during transport to the slide or if the
Pick&Place module
cannot remove a coverglass from the coverglass cartridge after 3 attempts, the user is notified of this
through an error message
.
1
Fig. 110
Note
Processing in the respective coverslip line is interrupted in order to prevent damage to the specimens.
Specimens may become dried out during the interruption.
L
If this message
is displayed, the user must first check whether a coverglass has
been lost during transport to the slide or whether the
Pick&Place module
was unable to take a
coverglass from the coverglass cartridge.
1. To do this, start by opening the hood of the instrument and checking whether or not a coverglass
has been lost while being transported to the slide.
2. To remove the slide, press the red bar to the far right
and hold it in this position.
3. Make sure that the shifter
does not obstruct the safe removal of the slide. Move
the shifter all the way forward or back if necessary.
4. Carefully remove the slide that has not yet been coverslipped from the instrument
and store it safely outside the instrument. After correcting the malfunction, the
slide can be coverslipped manually.