128
Version 2.1, Revision K
Cleaning and Maintenance
7
7.4.1 Daily cleaning and maintenance
A
Coverglass cartridge:
•
Check the coverglass cartridge for broken glass
replacing the coverglass cartridge)
•
Check that the coverglass is seated correctly in the coverglass cartridge
6.3.3 Checking and replacing the coverglass cartridge)
.
B
Waste tray:
•
Check the waste tray for broken glass
P. 86 – 6.3.4 Emptying the waste tray)
empty it if necessary.
C
Pick&Place module:
•
Check the skids, suction cups and coverglass sensor pin for residual mountant and
broken glass
P. 86 – 6.3.5 Inspect Pick&Place module)
and clean if necessary
P. 114 – 7.2.10 Cleaning the Pick&Place module)
•
Replace deformed and/or damaged suction cups
P. 114 – 7.2.11 Exchange suction
.
D
Alignment pins:
•
Check for residual mountant and clean if necessary
E
Coverslip lines
L1
and
L2
:
•
Check for surface-dried residual mountant and clean if necessary.
F
Reagent vessels in the load drawer:
•
Replace the reagent in the reagent vessel
P. 32 – 4.6 Refilling consumables)
.
G
Unload drawer:
•
Check if there are still racks in the unload drawer and remove
.
H
Surfaces of the instrument:
•
Check surfaces of the instrument for residual reagent in the area of the load drawer
and clean if necessary. For this purpose, commercial household cleaner can be used
P. 105 – 7.2.1 Exterior surfaces, varnished surfaces, instrument hood)
.
I
Prime bottle:
• Check the fill level of the prime bottle and dispose of the contents in accordance with the
laboratory specification if necessary.
J
Shifter and shifter tongue:
•
Check the shifter and shifter tongue
for contamination and dried out
mounting medium. Soak a lint-free cloth with xylene and remove any mounting medium
residue that may be present.