Malfunctions and Troubleshooting
8
8.4 Replacing main fuses
Warning
The instrument should be shut down using the
Main Switch
and disconnected from the power supply in
the case of a instrument failure. The main fuses can then be inspected.
A suitable flat-head screwdriver must be used to prevent damage to the fuse holders.
Beware of defective fuses! Potential risk of injury due to broken glass! Wear suitable protective
clothing (safety goggles, cut-resistant gloves
P. 18 – 3.1 Standard delivery – packing list)
).
1. To do so, open the hood and unscrew the two fuse holders
on the top of the right
cover with a flat-head screwdriver
2. Check the fuse being used for damage.
1
Fig. 126
3. Remove the defective fuse from the holder and replace it with a new fuse.
4. Installation occurs in the reverse sequence of operations.